My Trainer

LOGIQ S8 XDclear 2.0

Set up the device

  • To set up the position sensing apparatus,

    For accurate image registration, ensure that you connect the two probe position sensors to the probe. Match the numbers when you attach the sensors connectors to the sensor unit.

    1. Attach the receiver probe bracket to the probe. Ensure that it is attached securely.



  • 2. Fix the position sensor to the bracket.
  • 3. Connect the receiver cable from the Position Sensor to the front of the Ultrasound system.

    4. Trace back each cable physically to insure the cable in position 1 on the probe is connected to position 1 on the console and that the cable in position 2 on the probe is connected to position 2 on the console.

     

  • 5. Attach the transmitter cable to the front of the Ultrasound system where the transmitter label is shown.

    6. Now that all the cables are connected, press the V-Nav key. The light above the transmitter cable blinks until the position sensing system has been initialized by the system. Once initialization is complete, the light remains green.

  • 7. Place electromagnetic transmitter next to patient with front toward scanning area of interest.

    Place the transmitter such that the entire area of interest is in the front hemisphere of the transmitter. The transmitter needs to be placed such that the face of the transmitter opposite of the cable (the field side) faces the working area